The electric fan allows the exhaust of toxic vapours generated inside your safety cabinet. The vapours are purified by the active coal filter and rejected without danger in the atmosphere.
CONFORMITY
Electric fan in compliance with ISO 5801, 9614/1, AMCA 210-85; 0 marking.
Electric fan in compliance with ATEX norm on request. Please contact us.
ADVANTAGES
Adaptable to all of our cabinets.
No modification of building's structure required.
Simple and aesthetic fitting.
Easy connection between the cabinet and the vent box.
Connection by flexible tube Ø 100mm (in option).
TECHNICAL SPECIFICATIONS
Filtration Box Cdf - A Model (Not Shown On The Picture)
Built in steel 12/10th.
White epoxy paint RAL 9010.
Filtration Box CDF - P Model
Built in Cellular PVC.
White colour.
Each box contains 2 independent compartments:
Filtration Compartment:
Active coal filter type « candle » H150 X Ø 150mm (CDFF15 model) must be changed at least once a year.
Filter settled on the box by 3 lugs and airtightness insured thanks to a rubber seal.
Ventilation Compartment:
Electric engine with scroll in PVC:
- 220/240 volts – 50Hz – 52,3dB(A)
- Flow 80m3/hour - Protection IP24
- Exhaust intlet Ø 100 mm / outlet Ø 75 mm.
OPTIONS
Connecting KIT (KRC)
This kit allows to connect easily your vent box to your safety cabinet:
1 connection unit Ø 100mm(ref. CHJ) to be fixed on the cabinet air outlet.
Disclaimer: Loba Chemie Pvt. Ltd. makes every effort to ensure that the information displayed on this web site is accurate and complete, however it is not liable for any errors, inaccuracies or omissions. Majority of the information on lobachemie.com is liable to change without any intimation or notice.